GENEActiv Action

GENEActiv Action

Please note: if recording at a sampling frequency of 1KHz is not vital within your study we recommend GENEActiv Original.

As part of our raw data actigraphy range, GENEActiv Action offers a unique recording sampling frequency of up to 1,000Hz. It is specifically designed for applications in performance measures such as professional sports, bone health and high impact activities, which demand wider range acceleration measurement and a higher sampling frequency.

  • +/- 16g acceleration measurement at 7.8mg resolution
  • 1KHz recording sampling frequency
  • Over 12 hours of continuous recording
  • Waterproof & robust
  • Rechargeable battery
  • Open source data format

Enquire & order

  • Specification

    Please note: specification is identical to that for the GENEActiv Original, with the exception of those items highlighted in orange below.

    Physical Parameters
    Size 43mm x 40mm x 13mm
    Weight 16g (without strap)
    Main Housing Material PC/ABS (medical device grade)
    Light Guide Material PC (medical device grade)
    Data Contact Material Gold-plated
    Fixings 20mm heavy duty spring bar
    Strap PU resin
    Battery Type Rechargeable lithium polymer
    Environmental Protection
    Moisture Ingress Water-resistant to 10m (IP67 – 1m 24hrs)
    Material Ingress Dust tight (IP67)
    Operating Temperature 5 – 40 deg C
    Mechanical Impact 0.5m drop resistant
    Measurement Capabilities
    Memory 0.5Gb non-volatile
    Logging Frequencies Selectable 100 or 1000Hz
    Maximum Logging Periods 12 hours @ 1kHz
    Internal Clock
    Type Quartz Real Time Clock
    Frequency 32.7768kHz
    Accuracy +/- 20ppm (+/- 1.7s per day)
    Acceleration Measurements
    Sensor Type MEMS
    Range +/- 16g
    Resolution 12 bit (7.8 mg)
    Light Measurements
    Sensor Type Silicon photodiode
    Wavelength 400 to 1100 nm
    Range 0 – 3000 Lux typical
    Resolution 5 Lux typical
    Accuracy +/- 10% @ 1000 Lux calibration
    Event Logger
    Sensor Type Mechanical membrane switch
    Temperature Measurements
    Sensor Type Linear active thermister
    Range 0 to 60 deg C
    Resolution 0.25 deg C
    Accuracy +/- 1 deg C
    Measurement Frequency Every 30s minimum
    USB Connection
    Device USB 2.0 Full Speed
    Charge Cradle Format 4 unit cradle USB 2.0 High Speed
    Charge Time 90% @ 2 hours; 100% @ 3 hours
    Data Download Time Maximum 15 minutes for 4 concurrent units
  • Regulatory
    This device complies with part 15 of the FCC Rules.
    This product is compliant with the Directive 2004/108/EC; tested to BS EN 61000-6-1 :2007 and BS EN 61000-6-3 :2007.